HT-LA561 High-Content Oxygen Analyzer

HT-LA561 High-Content Oxygen Analyzer

Short Description:

 


Product Detail

Technical Specifications

Applications

Product Tags

HT-LA561 High-Content Oxygen Analyzer

 

Technical parameters

Measurement Principle: Ion current (limiting current)
Display Mode: 128×64 dot matrix OLED
Measurement Range: Type E: 10.00%~98.00% O2Type F: 10.00%~99.99% O2
Measurement Accuracy: ≤±1.2%FS
Resolution: 0.01%
Repeatability: ≤±1.0%FS
Response Time: T90≤30S
Output Port: 4-20mA.DC (Non-isolated output, load resistance1KΩ)0-5V.DC (Non-isolated output, load resistance>10KΩ)

2 ways of programmable alarm output (Dry contact type, maximum load 220V/2A)

Communication Mode: RS232 (default) or RS485
Power Supply: AC220V±10% 50/60Hz, power dissipation <10VA
Environment Temperature: -10℃~+50℃
Storage Environment Humidity: <90RH, non-condensing
Operation Environment Humidity: <100RH, non-condensing
Sample Gas Temperature: 0℃~50℃
Sampling Mode: Gas in
Gas Inlet Pressure: Stable pressure (control the pressure within the recommended limits to make sure the flow is within 400~600mL/min)
Gas Outlet Pressure: Free venting (under safe conditions)
Background Gas: N2 or other inert gas
External Size: 100mm×100mm×115mm (H×W×D)
Hole-Cutting Size: 92mm×92mm (H×W)
Sensor Service Life: >4 years (under normal conditions of use)
Gas Circuit Port: Φ6 plastic tube joint (sensor built in)
Instrument Weight: Net weight 0.7Kg
Installation Method: Embedded mounting

 

The instrument features

a.   The sensor is built in and is easy and convenient for user to install and use;
b.   The original imported limiting current sensor consumes no service life when the power is off, and has a long service life, rapid response speed, and high measurement accuracy, and is easy to maintain, etc.;
c.   The instrument allows multi-point fractional liner calibration, meeting the need for an accurate measurement of oxygen content among the full range.
d.   No calibration gas is needed and the instrument is not affected by the oxygen content in the operation environment;
e.   Altitude height measurement module can be expanded, preventing the impact of atmospheric pressure changes on the measurement accuracy;
f.    With temperature and flow compensation and sensor thermostatic control function, impact from sample gas flow and environment temperature changes on sensor measurement accuracy could be eliminated;
g.   Embedded mounting is easy and convenient to install and maintain.

  • Previous:
  • Next:

  • Measure Range:       0-10/100/1000ppm,0-1%O2

    Accuracy:            <=±5% in the range of 0-10ppm;s±3% in the range of 0-100ppm;s±2% in the range of 0-100pm;

    Display Resolution:    0.01ppm

    Response Time:       T90<=65s(0-1000ppm)

    Measure Medium:     H2, He, hydrocarbons, inert gas, mixed gas & acidic gases

    Stability:            Zero drift <=±1% of FS/7d; Span drift <=±1% of FS/7d

    Sample Gas Flow:     800-1000ml/min

    Repeatability:        <=±2.5% F.S(0-10ppm);<=±1.5% F.S(0-100ppm);<=±1% F.S(0-1000ppm);

    Sample Gas Pressure:  35KPa-210KPa

    RecoveringTime:      Reducing concentration in the air to 1ppm in 60mins

    Sensor Life:          24months at 25℃ and 1 atm.

    Sensor Principle:      Electrochemical Sensor

    Warranty:           12 months analyzer; 12 months sensor

    Weight:             4.5kg

    Appearance dimension: 255mm×250mm×160mm(L×W×H)

    It can be used to measure the trace oxygen concentration in the fields of nitrogen separation from air, chemical process, magnetic materials such as high temperature sintering furnace gas protection, the electronics industry protective gas, glass & building materials industries, various mixed gases, etc.