Varying levels of protection can be achieved by varying the thickness of lead shielding. Lead thickness is usually expressed in mm Pb.
The specification of effective lead shielding is determined by the level of protection required.The level of protection required is dictated by the equipment, room orientation, amount of use and activities planned for the installation.
Sunny Young provide solutions to the following international standards:
1、 Radiation Protection Institute of China, licensing requirements
2、 EU Medical Ionising Radiation Protection Regulations.
3、 National Council on Radiation Protection & Measurements (USA)
Work Procedure
Step 1 – Design Stage
Step 2 – Production Stage
Step3-Installation
Step 4-Commissioning
Mammogram
Material :
Lead plate thickness :1mm
Fixing steel frame :50*25 square steel tube;
Radiation-proof automatic translation door:
Size :1500*2200mm
Lead equivalent:2.0mmPb
Electrochemical steel plate coating baking paint
Installation:
1. Making the fixing steel frames, crossing type, horizontal and vertical spacing shall not exceed 60cm.
2. Lead plate installation and fixing;
3. The wall sharing with CT room will be blank.
4. Total area =Three walls=41㎡
5. The ceiling will be blank.
CT Scan
Material :
Walls Lead plate thickness :3mm
Ceiling Lead plate thickness :2mm
Fixing steel frame :50*25 square steel tube;
Radiation-proof automatic translation door:
Size :1500*2200mm
Lead equivalent:4.0mmPb
Electrochemical steel plate coating baking paint
Observation window- Lead Glass
Size :800*1200*20mm
Lead equivalent: 4.0mmPb
Including the stainless steel decorative frame